CT Number Conformity Test on Multislice CT Scan at Yogyakarta PDHI Islamic Hospital
Uji Kesesuaian CT Number pada CT Scan Multislice di Rumah Sakit Islam Yogyakarta PDHI
DOI:
https://doi.org/10.14421/physics.v6i1.5140Keywords:
Conformity test, CT Number, ROI, Water Phantom, IndoQCTAbstract
This research focuses on determining the CT number, which is the attenuation coefficient value of x-rays after passing through an organ. The level of energy attenuation depends on the initial energy of the x-rays and the atomic number of the object, which play a role in describing differences in the characteristics of organs or body tissues. This research aims to ensure the suitability of the CT number to obtain accurate information about CT scan images. Research procedures include observation, data collection, and data processing. The CT number data from the experimental results was then compared with the simulation results using IndoQCT software. The CT Number conformity test is carried out on a water phantom by adjusting the tube voltage, input current, and slice thickness. The Region of Interest (ROI) is made in the form of a circle at five measurement points, namely the central direction, 3 o'clock, 6 o'clock, 9 o'clock, and 12 o'clock. The research results, both direct testing and simulation using IndoQCT software, show that the CT number does not exceed the accuracy tolerance limit and uniformity, respectively, -4 ˂ HU ˂ +4 and -2 ˂ HU ˂ +2. Based on the analysis results, it can be concluded that the CT number measurements in the Radiology Unit at the Yogyakarta PDHI Islamic Hospital meet the requirements of BAPETEN Perka Number 2 of 2018.
References
M. Maqbol, An Introduction to Medical Physics, Cham, Switzerland: Springer, 2017.
A. Y. Nurhayati, N. N. Nariswari, B. Rahayuningsih, and Y. C. Hariadi, "Analisis Faktor Eksposi dan Ketebalan Irisan Terhadap CTDI dan Kualitas Citra Pada Computed Tomography Scan," BERKALA SAINTEK, vol. 7, no. 1, pp. 7–12, 2019.
K. L. Bontrager, Textbook of Radiographic Positioning and Related Anatomy, 8th ed., St. Louis: Mosby Inc., 2014.
E. Seeram, Computed Tomography: Physical Principles, Patient Care, Clinical Application, and Quality Control, 5th ed., Missouri: Elsevier Inc., 2023.
S. Puspaningrum, N. P. R. Jeniyanthi, and I. M. P. Darmita, "Analisis Nilai CT Number Pada Pemeriksaan CT Scan Thorax Pada Kasus Efusi Pleura di RS Bhayangkara Makassar," Nautical: Jurnal Ilmiah Multidisiplin, vol. 1, no. 10, 2023.
L. A. Nansih, R. A. Buretri, and C. Artitin, "Uji Kesesuaian CT Number in Water pada Pesawat CT Scan Merk Philips di Instalasi Radiologi RSU BMC Padang," Jurnal Teras Kesehatan, vol. 6, no. 1, pp. 16–21, 2023.
M. Zakirin, A. A. A. Diartama, M. Iffah, B. Mughie, and N. P. R. Jeniyanthi, "Uji Kesesuaian CT Number pada Pesawat CT Scan Multislice di Instalasi Radiologi RSUD Magunsada Badung," Imejing Jurnal, vol. 3, no. 1, 2019.
N. L. K. Sari, D. T. Bahagia, P. Haryanto, and D. Muliyati, "The Effects of High Dose and Low Dose Protocols in Thorax’s CT Scan Image Quality," Indonesian Journal of Applied Physics, vol. 11, no. 2, pp. 189–199, 2021.
P. W. Chu et al., "Reference Phantom Selection in Pediatric Computed Tomography Using Data from a Large, Multicenter Registry," Pediatric Radiology, vol. 52, no. 5, pp. 445–452, 2022.
N. N. Ahadiyah, N. Setiawati, Z. Arifin, and C. Anam, "Pengukuran Computed Tomography Dose Index (CTDI) pada Phantom Kepala dengan menggunakan CT Dose Profiler," Berkala Fisika, vol. 23, no. 4, pp. 118–125, 2020.
C. Anam, A. Naufal, W. S. Budi, H. Susanto, F. Haryanto, and G. Dougherty, Manual Book IndoQCT Version 22: Software for Evaluating the Quality of Computed Tomography Images, Semarang: Universitas Diponegoro, 2020.
BAPETEN, Peraturan Kepala BAPETEN No. 2 Tahun 2018 tentang Uji Kesesuaian Pesawat Sinar-X Radiologi Diagnostik dan Intervensional, 2018.
R. A. Schulz, J. A. Stein, and N. J. Pelc, "How CT Happened: The Early Development of Medical Computed Tomography," Journal of Medical Imaging, vol. 8, no. 5, 2021.
N. N. Ratini, G. N. Sutapa, N. K. N. Anggarani, and A. J. G. Mberato, "Measurement of CT Number Uniformity Value at Various X-ray Tube Current Settings in CT Scanning," Jurnal Penelitian Pendidikan IPA, vol. 11, no. 1, pp. 904–909, 2025.
A. J. G. Mberato, N. N. Ratini, I. R. Yuliara, and R. Irhas, "Analysis of the Effect of the X-ray Tube Current on Uniformity Image Noise Value," Indonesian Physical Review, vol. 6, no. 1, pp. 11–19, 2023.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2025 Ali Mas'uul

This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.